XAFS

  This equipment is designed for the XAFS in transmission and fluorescence modes. The energy range of 4.6 - 70 keV is possible to be utilized. To keep the XAFS data with high precision, double-mirror configuration in transport channel is adopted combined with a double-crystal monochromator. XAFS measurements in the transmission mode can be performed using ionization chambers and detection gases suitable for the photon energy. The Lytle-type detector and the 19-element Ge solid-state detector are installed for the fluorescence mode measurement. The cryostat system for adjustment of the sample temperature and a high-precision θ-2θ stage for glancing-angle XAFS measurements are also furnished.

XAFS_image.jpg

Schematic view of XAFS measurement system

  • Facilities
  • Detectors
    • Ionization chamber
    • 19-element Ge solid-state detector with 100 mm2 x 10 mm-thick. x 19 elements
    • Lytle-type detector
    • Detection system for conversion-electron yield
  • Cryostat system, control range of sample temperature
    • about 9 k through 300 k with stability of 0.1 k
  • Disposal system for exhausting gases (the system will be utilized for the catalysis research)