大型放射光施設 SPring-8

コンテンツへジャンプする
» ENGLISH
パーソナルツール
 

BL22XU HAXPES

問い合わせ番号

INS-0000001607

HAXPES

 Utilizing a large probing depth of high kinetic energy photoelectrons, the surface contamination effect is drastically reduced. Especially, it is suitable for analysis of samples that have difficulty in surface cleaning. Moreover, this method is powerful to clarify the chemical states of environmental samples.

 The system consists of a SCIENTA-OMICRON WE4000 large-acceptance-angle spectrometer combined with double crystal monochromator at BL22XU. Angle dependence of PES in hard x-ray region can also be performed using this system, which provides the electronic states from bulk to surface. The Combined low-energy electron / ion flood gun is installed to measure insulation. The system uses Kirkpatrick–Baez focusing mirrors to achieve a beam size of 2.3μm at the photon energy of 8keV. Through a combination of K-B mirrors and sample scanning, chemical state mapping is possible.

Fig. 1  Hard x-ray photoelectron spectroscopy system

Fig. 1 Hard x-ray photoelectron spectroscopy system

最終変更日