大型放射光施設 SPring-8

コンテンツへジャンプする
» ENGLISH
パーソナルツール
 

BL43IR Infrared microspectroscopy station

問い合わせ番号

INS-0000001428

Infrared microspectroscopy station

  Our infrared microscope at this station is capable of microanalysis of various specimens as small as 10 µm diameter with the infrared synchrotron radiation that is highly collimated and highly brighter than the conventional infrared light source.

  1. Wide wavenumber region (100 ∼ 20000 cm-1).
  2. Polarizers cover the entire wavenumber region.
  3. Spatial resolutions are 10 µm without an aperture, 5 µm with an aperture in the mid-infrared region. In the far-infrared region, the resolution reaches the diffraction limit.
  4. Transmission and reflection measurements.
  5. Long working distance (100 mm (± 50 mm)) between Schwarzschild mirrors ( × 8, NA = 0.5).
  6. In-situ measurement of ruby fluorescence for the DAC study.
  7. All optical paths are possible to be purged with the dry air.
  • Sample stages and accessories
    • X-Y-Z motor-controlled stage
      • Working length : X-Y: 100 mm, Z: 10 mm, Minimum step: 1 µm
    • Bruker X-Y mapping stage
      • Working length : X-Y: 100 mm, Minimum step: 1 µm
    • Flow-type cryostat
      • Microstat-He : Oxford Instruments Co. Ltd.
      • Temperature : 4.2 ∼ 400 K
      • Window : Quartz, KRS-5, BaF2, Polyethylene
    • Low temperature DAC
      • Gas-membrane type DAC : ∼ 20 GPa @ 0.6 mm curette
      • Temperature : 10 ∼ 400 K
      • Window : Quartz, KRS-5, BaF2, Polyethylene
    • High temperature DAC
      • Lever type, external heating HTDAC (∼ 700°C, ∼ 30 GPa)
最終変更日